Dual Source Scatter Phantom


Price: Call JRT for Pricing
Item Number: 76-840



The Dual Source scatter phantom simulates in-vivo forward and backscatter characteristics of 99mTc gamma rays for the extrinsic measurement of a scintillation camera’s deadtime. The phantom produces a spectrum typical of that observed from 99mTc in the myocardium.

Reference: Ralph Adams, Gerald J. Hine, and C. Duane Zimmerman, “Deadtime Measurements in Scintillation Cameras Under Scatter Conditions Simulating Quantitative Nuclear Cardiography,” The Journal of Nuclear Medicine, 19 (1978), 538-544.

  • Made of PMMA
  • Dimensions: 150 x 200 x 200 mm
  • Two holes are used to hold the radioactive sources
    • Hole dimensions: Ø 17mm x 120mm deep
    • Spaced 50 mm apart (center-to-center) 
    • Distance from the face of the phantom: 50 mm
  • Complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • NEMA Standards Publication (NU-1 2007) Gamma Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified