Pro-NM Performance Small
|
|
Price: Call JRT for Pricing
|
The Pro-NM Performance SMALL is used for small FOV NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.
Technical data (can be modified to customer specifications):
- Main cylinder:
- Inside cylinder diameter: 140 mm
- Inside cylinder height: 150 mm
- Cylinder wall thickness: 6 mm
- Cold rods insert:
- Rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1mm – height of rods: 40 mm
- Cold spheres:
- Solid sphere diameters: 3.2, 4.8, 9.5, 12.7, 15.9 and 19.1mm – height of the center of the spheres from the base plate: 78.5 mm
- Optional PET Lid with cylindrical samples:
- Refillable thin-walled cylinders, diameters: 4, 6, 8, 12, 16mm
- PTFE solid cylinder diameter: 16 mm
- Cylinder height: 25.4 mm
- Optional heavy duty carrying case
The Pro-NM Performance SMALL complies with:
- NEMA – Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
| 08-111
|
Pro-NM Performance SMALL
|
| 08-112
|
Pro-NM Performance SMALL with PET Lid
|
|
|
|