The Pro-NM Performance MINI is used for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45mm or 75 mm.

Technical data (can be modified to customer specifications):
  • At least 75mm FOV version:
    • Main cylinder:
      • Outside cylinder diameter: 80 mm
      • Inside cylinder diameter: 74 mm
      • Inside cylinder height: 65 mm
    • Cold rods insert:
      • Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
      • Height of rods: 34 mm
  • At least 45mm FOV version:
    • Main cylinder:
      • Outside cylinder diameter: 50 mm
      • Inside cylinder diameter: 45 mm
    • Inside cylinder height: 65 mm
    • Cold rods insert:
      • Rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
      • Height of rods: 34 mm
  • Optional carrying case
The Pro-NM Performance MINI complies with:
  • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
  • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
  • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
  • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals