NEED HELP?
Call JRT at 800-221-0111
8 am to 5 pm EST
|
Pro-NM Performance HD
|
|
Price: Call JRT for Pricing
|
The Pro-NM Performance HD is used for high resolution NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter
compensation.
Technical data (can be modified to customer specifications):
- Main cylinder:
- Inside cylinder diameter: 206 mm
- Inside cylinder height: 186 mm
- Cylinder wall thickness: 7 mm
- Cold rods insert:
- Rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1 mm
- Height of rods: 88 mm
- Cold spheres:
- Cold sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm
- Height of the center of the spheres from the base plate: 127 mm
- Optional PET Lid with cylindrical samples:
- Refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm
- Water filled cylinder diameter: 25 mm
- Air filled cylinder diameter: 25 mm
- PTFE solid cylinder diameter: 25 mm
- Cylinder height: 38 mm
- Optional heavy duty carrying case
Pro-NM Performance HD complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
| 08-106
|
Pro-NM Performance HD
|
| 08-107
|
Pro-NM Performance HD with PET Lid
|
|
|
|