Browse Categories
 
         


 
 
   NEED HELP?
                
 
    Call JRT at 800-221-0111 
           8 am to 5 pm EST
 
 
 
 
 

Pro-NM NEMA NU2 Resolution


Price: Call JRT for Pricing
Item Number: 08-109


The Pro-NM NEMA NU 2 Resolution Phantom is used for evaluation of spatial resolution of positron emission tomographs (PET). It is used to characterize the widths of the reconstructed image point spread functions (PSF) of compact radioactive sources. It has been designed in accordance with the NEMA Nu 2-2012.

The Pro-NM NEMA NU 2 Resolution Phantom is compatible with Diagnomatic QC Solutions, a web-based service that makes it possible to easily organize and manage the quality control of any number of diagnostic imaging systems. The platform supports a DICOM protocol that allows seamless integration with your hospital IT system making data transfer safe and easy. The Pro-NM PET BASIC Kit is a versatile set of phantoms and software for carrying out constancy tests of radiography and fluoroscopy units. Thanks to the Diagnomatic subscription, all tests can be quickly and effortlessly automatically analyzed online and with the desktop app.

Technical data (can be modified to customer specifications):
  • Five capillaries with an inside diameter of 1 mm and an outside diameter of less than 2 mm
  • In the transverse direction the capillary is positioned at 1 cm
  • (to represent the center of the FOV, but positioned to avoid any possible inconsistent results at the very center of the FOV), 10 cm, and 20 cm from the center of the plane
  • Middle of the capillaries is marked to provide guidance while the radionuclide is introduced
  • Markers and convenient base with a spirit level for easy and accurate positioning
  • Optional carrying case
The Pro-NM NEMA NU 2 Resolution Phantom complies with:
  • International Standard: Radionuclide imaging devices – Characteristics and test conditions – Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC),61675-1, Geneva, Switzerland, 1998.
  • Performance Measurements of Positron Emission Tomographs, NEMA Standards Publication No. NU2-2012, National Electrical Manufacturers Association (NEMA), 2012.
  • NEMA2007/IEC2008
  • NEMA 2012/IEC 2008