Doppler Flow Test System

The Doppler Flow Test System 1425A LE tests both Doppler and B-mode ultrasound systems using one uniquely designed unit. A low echo matrix is combined with strategically placed line reflectors and anechoic cyst targets at 2, 4, and 6 mm depths.

Two 5 mm vessels are incorporated into the system to meet with FDA Doppler sensitivity recommendations. One vessel is parallel to the scan plan at 2 cm, replicating a carotid artery. The other vessel descends 45 degrees from the surface to test Doppler sensitivity and to train different scanner techniques. All of this is combined with a flow controller with microprocessor that has a range of 1 to 12.5 ml/sec within 3% FS accuracy.

You can access five programmable test programs and five preset pulse flow patterns with the easy-to-use display panel. This allows for accurate reproducible testing time after time.

By combining the flow system, phantom, and electronic flow controller into a self-contained dual purpose Doppler Flow Test System, the 1425A LE is an excellent instrument for performing a multitude of functions. Scanner selection, quality control testing, training, and research can all be performed using this multi-faceted ultrasound tool.

All of the following quality indicators can be measured in real time:

  • Maximum signal penetration
  • Channel isolation or directional discrimination
  • Registration accuracy of duplex sample gates and similarities between B-mode and color flow images
  • Flow rate readout accuracy for various angles, beam directions and operating modes

The Doppler Flow Test System 1425A LE is not only one of the most versatile dual-purpose systems available, but it’s completely portable as well. This makes our Doppler Flow System a must-have for your ultrasound quality control.

1425A-LE-0.5 Doppler Flow Test System, 0.5 Attenuation

1425A-LE-0.7 Doppler Flow Test System, 0.7 Attenuation

 

Doppler Flow Test System

 

SPECIFICATIONS

Tissue Mimicking Material
Speed of sound
1540 ±10 m/s Attenuation 0.5 or 0.7 ±0.05 dB/cm/MHz

Targets
Cyst speed of sound
1540 ±10 m/s Cyst attenuation 0.05 ±0.01 dB/cm/MHz
Diameter of nylon lines 0.1 mm (0.004 in)
Size tolerance ±5% Location tolerance . . . . . . ±0.1 mm (0.004 in)

Vessels
Size
5 mm diameter
Location 2 cm below surface; 2 to 16 cm at 40°angle

Blood Mimicking Fluid
Density
1.03 g/cc
Speed of sound 1550 ±10 m/sec Scatter size 4.7 micrometers average

Electronic Flow Control System
Constant flow mode
1 to 12.5 ml/sec ±3%
Calculated velocities 10-110 cm/sec Pulsatile flow mode 5 programmed waveforms
Container Size 40.6 x 24.8 x 40 cm (16 x 9.75 x 15.75 in)
Weight 10.4 kg (23 lbs)

All acoustic measurements at 4.5 MHz, 22°C